The AQUALYTIC® COD vario test setup allows precise water testing with minimum effort. After adding the sample to a AQUALYTIC® COD vario tube test (LR, MR according to ISO 15705:2003-01) it is heated in the thermoreactor AL125 and then analyzed photometric.
The COD- Setup comprises the AL250 COD vario photometer, 25 tube tests for the measuring ranges 0-150 und 0-1500 mg/l O2, a thermoreactor AL125 for sample digestion, and a vial stand.
With 3 measuring ranges from 0 bis 15000 mg/l O2, the AQUALYTIC® AL250 COD vario photometer is suitable for waste water testing. Tow LEDs light sources with long-term stability, a waterproof sample chamber, a large digital display, and the user-friendly keypad ensure maximum operating reliability and convenience.
Technical data: |
Meas. cycle: | approx. 3 seconds |
Optics: | temperature-compensated LEDs (λ1= 605 nm, λ2 = 430 nm), and photo-sensor amplifier in waterproof sample chamber |
Power supply: | 9V power pack battery providing 40 hours operating |
Auto - OFF: | automatic switch-off approx. 5 minutes after last key press |
Ranges: | 0 – 150 mg/l O2 ±3.5 %* 0 – 1500 mg/l O2 ±3.5 %* 0 – 15000 mg/l O2 ±3.5 %* *based on potassium-hydrogen phthalate standards (DIN 38409) |
Test instrument suitability: | Calibration and software based adjustment options. |
CE-conformity: | DIN EN 50081-1 - VDE 0839 part 81-1:1993-03 DIN EN 50082-2 - VDE 0839 part 82-2:1996-02 |
Use the membrane keyboard to set the temperature and choose the reaction time. The reactor comprises three temperature ranges (100 / 120 / 150 °C) and three pre-programmed reaction times (30 / 60 / 120 min.). After completion of the digestion, the reactor switches off automatically, indicated by an acoustic alarm and an LED signal.
The reactor AL125 has 24 holes to receive 16 mm round tubes.