Chemic ALS - Via S.Francesco d'Assisi, 200 - 70018 Rutigliano (BA) - Italy - tel. (+39)080.476.10.66 - fax (+39)080.476.11.26 - info@labcatalogo.it - www.labcatalogo.it
Main -> Laboratory facilities -> Protective equipment -> Fume hoods
Fume Hoods, Köttermann
Fume cupboards for general use

The laboratory hood is the type of hood which is typically used in laboratories. It can be used for all tasks in the laboratory.

Equipment:
Interior height 1500 mm
One-part front slide with sliding windows made of laminated safety glass
AutoProtect
Ceramics counter top with funnel sink, both blue-grey
Digital air stream monitoring with pressure transmitter
Interior lighting
Media equipment:
4 plugs 230 V/16 A, German
1 × cold water
Ventilation certified according to DIN 12924 part 1

W × H × D
mm
EquipmentOrder No.
1200×2550×9002 sliding panels10 12 01005
1500×2550×9003 sliding panels10 12 01006
(1)1800×2550×9003 sliding panels10 12 01007
2100×2550×9004 sliding panels10 12 01008



Without the pictured accessories.

High performance fume cupboards

Fume hoods - more precise: hoods for working with aggressive media at high temperatures - are used when greater heat volumes are released, e.g. when fuming acids, or when working with aggressive chemicals.

Equipment:

  • Interior height 1200 mm
  • Digital air stream monitoring with pressure transmitter
  • Interior lighting
  • Media equipment: 4 plugs 230 V/16 A, German
    1 × cold water
    1 × burnable gas
  • Ventilation certified according to DIN 12926 part 2 and EN 14175-3
  • For connection to available venting system
  • Including AutoProtect, the intelligent front sliding controller

Variants
Fume hoods, ceramics

  • One-part front-slider with fixed window made of laminated safety glass
  • Ceramics counter top with funnel sink, both blue-grey
  • Interior lining made of extensive white ceramics
W × H × D
mm
Order No.
1200×2550×90010 12 01017
1500×2550×90010 12 01018
(2)1800×2550×90010 12 01019


High performance fume cupboards, Polypropylene

Especially for working with hydrofluoric acid

  • One-part front slider with fixed window made of polycarbonate
  • Polypropylene counter top with funnel sink, both blue-grey
  • Interior lining made of extensive white polypropylene
W × H × D
mm
Order No.
1200×2550×90010 12 01021
1500×2550×90010 12 01022
1800×2550×90010 12 01023


Pharmacy fume cupboards

With their certification according to DIN 12924 part 4, pharmacy hoods comply with the requirements of pharmacy regulations. They are perfectly suitable for handling dusty preparations or volatile chemicals.

Equipment:

  • Hood for setting up on a table in sitting work height (750 mm)
  • Self-supporting housing made of galvanised and epoxy-coated steel
  • Colour: light-blue (similar to RAL 7035)
  • Certified according to DIN 12924 part 4
  • GS-certificate by TÜV Product Service
  • Air stream monitoring unit according to DIN 12924, battery-buffered
  • Front slider made of polyacrylics
  • Counter top made of epoxy-coated steel
  • Power supply: 230 V
  • Exhaust connection: DN 110
  • Work space lighting
  • Media equipment:
    1 × burnable gas
    2 × 230 V
    1 × water
W × H × D
mm
EquipmentOrder No.
900×1350×640integrated fan10 12 01001
(3)900×1350×640without fan10 12 01002



Underbench cabinet optional

Extractor hoods

Table exhaust hoods are used to exhaust minor dangerous pollutants, which could be released for example during refilling. Because they are covered from three sides, they provide a quite good exhaust performance. Table exhaust hoods are not a replacement for laboratory hoods, because they are not ventilation-certified.

Equipment:

  • For setting up on a table in sitting or standing working height
  • Self-supporting housing made of galvanised and epoxy-coated steel
  • Colour: light-blue (similar to RAL 7035)
  • Front slider made of polyacrylics
  • Exhaust connection: DN 110
  • Work space lighting
W × H × D
mm
EquipmentOrder No.
800×1350×640integrated fan10 12 01003
(4)800×1350×640without fan10 12 01004



Underbench cabinet optional

Directed extractions
ArticleOrder No.
(5)Directed extraction for wall fastening, DN 7510 12 01011
(6)Directed extraction for ceiling suspension, DN 7510 12 01012


Extraction flues
ArticleW × H × D
mm
Order No.
(7)Extraction flue, DN 2001175×500×60010 12 01031
Extraction flue, DN 2001175×500×75010 12 01032



The pictured hoods are planning examples. We would be glad to compose an individually equipped hood according to your requirements. Cabinets, safety cabinets, chemical cabinets or acid-leach cabinets are not included in the scope of delivery. They can be ordered as accessories and can be upgraded without any problems.