Chemic ALS - Via S.Francesco d'Assisi, 200 - 70018 Rutigliano (BA) - Italy - tel. (+39)080.476.10.66 - fax (+39)080.476.11.26 - info@labcatalogo.it - www.labcatalogo.it
Main -> Measuring systems -> Electrochemistry -> pH multi-parameter measuring instruments
Multi Parameter Instruments, Handhelds, WTW

Multi 350i
Multi-parameter measurement devices in service-proven WTW handheld instrument design, dust-proof and water-proof, for simultaneous measurement, indication and storage of up to 4 parameters: pH/mV/ISE, dissolved oxygen, conductivity and temperature. Menu-controlled with a backlit graphic display. Data logger and memory for more than 1800 entries, GLP-supporting functions, real-time clock, calibrating timer. Bidirectional RS 232 interface. Suitable for connecting a large selection of sensors.
Complete with NiMH rechargeable batteries, wide-band mains power supply and KOM pilot software package for data collection on a PC.

ConOx combined conductivity/oxygen sensor
Simultaneous measurement of dissolved oxygen, conductivity and temperature with just one sensor. The ideal partner in water analysis for the Multi 350i. Automatic salinity correction, modular design. Conductivity module with service-proven 4-pole technology, galvanic oxygen module.

Multi-parameter sensor MPP 350
One sensor, 4 parameters: pH, conductivity, dissolved oxygen, temperature. With cable lengths up to 25 m, excellently suited for parallel field measurements in environmental analysis. The sensor is perfectly tuned to the Multi 350i. Complete with accessories, without pH electrode (must be ordered separately).

 

pH/mV/ISE measurement

Measurng ranges:

-2.000 to 20.000 pH
-2000 to 2000 mV
-999.9 to 999.9 mV
0.000 to 2000 mg/l in 4 ranges

Accuracy:

±0.004 pH, ± 1 mV, ± 0.2 mV

Temp. compensation:

automatic -5 to 105.0°C
manual -20 to +130°C

Oxygen measurement

Measuring ranges:

O 2 conc. 0.00 to 20.00 mg/l, 0.0 to 90.0 mg/l
O 2 sat. 0.0 to 200.0%, 0 to 600%

Accuracy:

±0.5 % of measured value

Temp. compensation:

automatic IMT compensation 0.0 to 50°C

Conductivity measurement

Measuring ranges:

0.001 to 1.999 µS/cm; K=0.01 cm-1
0.01 to 19.99 µS/cm; K=0.1 cm-1
0.1 µS/cm to 2000 mS in 5 ranges
Salinity 0.0 to 70.0
TDS 0.0 to 2000 mg/l
Tref 20°C or 25°C switch-selectable
Temp. -5.0 to 105.0°C

Accuracy:

±0.5 % of measured value

Temp. compensation:

nlF, linear 0.001 to 3.000 %/K, can also be switched off

ConOx data:

Shaft length/diameter:

approx. 145 mm/15.3 mm

Cell constant:

0.475 cm-1

Measuring range:

1 µS/cm to 2 S/cm

Temp. range:

0 to 50°C

Typ O 2 sensor:

galvanic



ArticleOrder No.
Multi-parameter meter Multi 350i (single unit)61 81 40110
(1)Multi-parameter meter Multi 350i (set)61 81 40114

with ConOx-3 m, SenTix 41-3 and accessories in the professional case

Multi-parameter meter Multi 350i (set)61 81 40115

MPP 350-6, incl. pH electrode, accessories in carry case

Multi-parameter meter Multi 350i (set)61 81 40116

with MPP 350-25, incl. pH electrode, accessories in carry case

SensoLyt MPP-A, reinforced pH electrode for MPP 35061 81 40152
NT multirange wide-band power supply unit61 81 02867