Chemic ALS - Via S.Francesco d'Assisi, 200 - 70018 Rutigliano (BA) - Italy - tel. (+39)080.476.10.66 - fax (+39)080.476.11.26 - info@labcatalogo.it - www.labcatalogo.it
Main -> Measuring systems -> Electrochemistry -> pH electrodes, buffer solutions, accessories
pH Electrodes, METTLER TOLEDO
Routine pH electrodes

These InLab® electrodes cover most laboratory applications in aqueous samples. Due to ARGENTHAL®-lead off system and silver ion scavenger, a contamination by sulfide-containing media is excluded and an improved temperature stability can be attained. This ensures that the electrolyte is completely free of silver ions and therefore also suitable for biological specimens.

 

Technical data:

Shaft material:

glass

Reference electrolyte/storage:

3 mol/L KCl

Temperature range:

0 – 100°C

Shaft diameter:

12 mm

Reference:

ARGENTHAL® with Ag-scavenger



TypepHTemp. sensorInstalled length
mm
Diaphragm modelConnectionOrder No.
(1)Routine0 – 14120CeramicS710 50 03050
(2)Routine-L0 – 14170CeramicS710 50 03053
(3)Routine Pro0 – 14NTC 30kΩ120CeramicMultiPin10 50 03054
(4)Routine Pt10000 – 14Pt1000120CeramicMultiPin10 50 03056
(5)Routine Pro-L0 – 14Pt1000170CeramicDIN cable/4 mm10 50 03057
(6)Science0 – 12120flex. glass grindingS710 50 03070
(7)Science Pro0 – 12NTC 30kΩ170flex. glass grindingMultiPin10 50 03071


Professional pH electrodes

With their open connection, the electrodes are especially unsusceptible to contaminations. The polymer reference electrode on these sensors is in direct contact with the measurement medium, therefore the diaphragm cannot clog. The PEEK-shaft is extremely resistant both chemically and mechanically.

 

Technical data:

Shaft material:

PEEK (state-of-the-art)

Reference electrolyte:

polymer electrolyte XEROLYT®

Temperature range:

0 – 100°C

Shaft diameter:

12 mm

Diaphragm type:

open connection

Reference system:

ARGENTHAL®

Storage:

FRISCOLYT-B®



TypepHTemp. sensorInstalled length
mm
Connec.Order No.
(8)Expert0 – 14120S710 50 03100
(9)Expert Pro0 – 14NTC 30 kΩ1201.2 m BNC cable10 50 03101
(10)Expert DIN0 – 14Pt10001201.2 m DIN cable10 50 03103
(11)Expert NTC300 – 14NTC 30 kΩ120MultiPin10 50 03104
(12)Expert Pt10000 – 14Pt1000120MultiPin10 50 03105


pH expert electrodes

Extreme specialist or top generalist - these electrodes are either one or the other. Here, you can find special electrodes for small, cold samples containing hydrofluoric acid and pure, viscous and very adherent samples. The puncture electrode InLab®Solids was designed for ph-measurement for foods such as fruit, cheese or meat. The InLab®Surface was developed for ph-measurement of surfaces, e.g. paper or skin. It is also perfectly suitable for determining drop-sized samples on a glass carrier.

TypepHTemp. sensorTemp. range
°C
Shaft (d / l)
mm
Connec.Order No.
(13)Solids1)1 – 110 – 806/25S710 50 03153
(14)Solids Pro1)1 – 11NTC 30 kΩ0 – 806/25MultiPin10 50 03154
(15)Viscous2)0 – 140 – 1006/40S710 50 03150
(16)Reach3)0 – 140 – 10012/300S710 50 03060
(17)Reach Pro3)0 – 14NTC 30 kΩ0 – 10012/400MultiPin10 50 03061
(18)Reach Pt10003)0 – 14Pt10000 – 10012/400MultiPin10 50 03062
(19)Surface5)1 – 110 – 5012/110S710 50 03157
(20)Semi-Micro6)0 – 120 – 1006/100S710 50 03165
(21)Semi-Micro-L3)0 – 140 – 1006/230S710 50 03161
(22)Micro3)0 – 140 – 803/60S710 50 03160
(23)Micro Pro3)0 – 14NTC 30 kΩ0 – 1005/130MultiPin10 50 03162
(24)Cool4)1 – 11-30 – 8012/120S710 50 03174
(25)Hydrofluoric3)1 – 110 – 10012/120S710 50 03176
(26)Pure7)1 – 110 – 8012/120S710 50 03170
(27)Pure Pro3)1 – 11NTC 30 kΩ0 – 8012/170MultiPin10 50 03171
(28)Power2)0 – 120 – 10012/120S710 50 03110
(29)Power Pro2)0 – 12NTC 30 kΩ0 – 10012/170MultiPin10 50 03111



Reference system/ reference electrolyte:
1) ARGENTHAL / XEROLYT®PLUS polymer
2) SteadyForce / FRISCOLYT-C®
3) ARGENTHAL with Ag-scavenger / 3 mol/KCl
4) ARGENTHAL with Ag-scavenger / FRISCOLYT-B
5) Ag/AgCl / 3 mol/KCl AgCl saturated
6) ARGENTHAL / XEROLYT®EXTRA polymer
7) Ag/AgCl / gel

Economy ph electrodes

The economy ph-electrodes offer en excellent price-performance ratio. With maintenance-free gel electrolytes, the InLab®Easy Din, Easy BNC and Easy are easy to operate and thus are perfectly suitable for certain applications, e.g. in schools.

TypepHTemp. sensorInstalled length
mm
Diaphragm modelConnectionOrder No.
(30)Easy0 – 14120CeramicS710 50 03010
(31)Easy BNC0 – 14120Ceramic1.2 m BNC cable10 50 03011
(32)Easy DIN0 – 14120Ceramic1.2 m DIN cable10 50 03012
Easy Pt10000 – 14Pt1000120Ceramic1.2 m DIN cable10 50 03015
(33)Basics BNC0 – 14120Ceramic1.2 m BNC cable10 50 03020
(34)Basics DIN0 – 14120Ceramic1.2 m DIN cable10 50 03021
Basics Pt10000 – 14Pt1000120Ceramic1.2 m DIN cable10 50 03023


35 PerfectION combined ion-selective electrodes

New sensors with improved measuring performance and optimized handling for a wide range of applications:

  • Click&Clear system for ultra-easy diaphragm cleaning and drift reduction
  • Plug&Measure kit incl. solutions, detailed application support, and background information
  • Polymer electrodes with replaceable membrane
  • Matching calibration standards can be ordered to go with the electrodes

Shaft material:
Epoxy (except for Na+: glass)

Scope of delivery:
Electrode with fixed cable BNC connector (DIN and LEMO connectors available on request), reference and ISA solution, guide book and information on the application

TypeMeasuring range
mg/l
Temp.
°C
Membraneoptimum pH rangeOrder No.
Ag+/S2-0.01 – 1080000 – 80Solid state2 – 1210 50 04700
Ca2+0.02 – 401000 – 40Polymer2.5 – 1110 50 04703
C1-1.8 – 355000 – 80Solid state2 – 1210 50 04706
CN-0.2 – 2600 – 80Solid state10 – 1410 50 04709
Cu2+0.00064 – 63540 – 80Solid state2 – 1210 50 04712
F-0.02 – gesättigt0 – 80Solid state4.5 – 5.510 50 04715
I-0.005 – 1270000 – 80Solid state0 – 1210 50 04718
K+0.04 – 390000 – 40Polymer2.5 – 1110 50 04721
NO3-0.1 – 140000 – 40Polymer2.5 – 1110 50 04727
Pb2+0.2 – 207000 – 80Solid state4 – 710 50 04730
Na+0.002 – 230000 – 80Na+ glass8 – 1110 50 04724