Chemic ALS - Via S.Francesco d'Assisi, 200 - 70018 Rutigliano (BA) - Italy - tel. (+39)080.476.10.66 - fax (+39)080.476.11.26 - info@labcatalogo.it - www.labcatalogo.it
Main -> Optical investigations -> Spectroscopy -> Photometers
Jenway Live Science Spectrophotometer Genova, Landgraf
Spectral photometer for measuring RNA/DNA samples and for protein determination

The Genova Photometer has been developed for scientists who need fast quantitative results for their DNA and protein sampling. With its integrated scanner, it is ideal for quality determination of oligonucleotides, DNA or RNA. Its software provides you with immediate quantification of ssDNA, dsDNA, RNA and oligonucleotides. You can also gauge the quality of your sample using the Genova's ratio calculation at 260/280 and 260/230 and the purity scan mode.
For the quantification of proteins, the Genova has a variety of measuring systems, such as Bradford, Lowry, Biuret, Biccinchonic and Direct UV Calculation.
The Genova Photometer is fitted with a Xenon flashlight. This lamp has a light intensity which is 5 to 10 times greater than for normal UV equipment. As a result it has a very good signal/noise ratio, particularly for wavelengths in biological samples. The Genova calibrates the wavelength independently and therefore satisfies GLP procedures.
In addition, the photometer „Genova Nano“ is equipped with a measuring cell with which samples of a volume of up to 2 μl can be measured. A cuvette is not needed. It is sufficient to pipette the sample on the measuring cell and to close it. After measuring the sample, it can be pipetted off the glass surface and the surface can be cleaned with a laboratory cloth. Due to the small layer thickness of 0.2 or 1 mm of the sample, the often necessary thinning of the sample is unnecessary.

 

Technical data:

Cuvette holder:

simple

Wavelength range:

198 – 1000 nm

Resolution:

1 nm

Accuracy:

±2 nm

Band width:

5 nm typically 270 nm

Resolution:

0.1%T, 0.001 A, 0.001 C

Concentration units:

ppm, mg/l, g/l, M,%, mg/ml, µg/ml, ng/ml

Photometric noise levels:

< 0.001A @ 0A @ 400 nm

Photometric stability:

<0.002 A/h after 30 minutes

Output:

Analogue output and RS 232 for computer and printer

Dimensions:

365 × 272 × 160 mm

Weight:

6 kg



ArticleVoltage
V
Connected load
Hz
Order No.
Genova live science photometer23050/6063 67 36001
Genova nano photometer23050/6010 67 26002


Accessories
ArticleOrder No.
Printer with printer cable63 67 50900
Cover hood63 67 30028