Standard scope of delivery
Venting and control systems:
- mandatory interior air circulation by means of circulation ventilators
- adaptive, fuzzy-controlled, multi-functional and digital PID-micro-processor control, for adjustable control of the Peltier heating/cooling system without cooling agents
- self-diagnostics system to find errors for temperature and humidity control
- 2 Pt100-sensors DIN-class A in 4-conductor measurement with changing monitoring and function adaption at exactly the same temperature
- digital real-time week-programme clock with day-group function, can be adjusted by the minute
- relative-time countdown clock for tempering profiles with up to 40 ramps, each segment can be set from 1 min to 999 hrs.
- digital display of all temperature, weekday, time, humidity and set-up default values - language selection via set-up
- digital target value specifications / current value display: 0.1°C
- integrated GLP-conform long-term logging (ring buffer) of all relevant data as data logging - 1024 kB
- saving the programme run in case of power failure
- parallel printer interface (incl. real-time clock with date function) for all PCL3-compatible ink jet printers for GLP-conform documentation
- USB interface including Memmert programming and logging software "Celsius" with advanced features
- chip-card control including 1 MEMoryCard XL with 32 kB memory (max. 40 ramps)
- incl. factory calibration certificate for +10°C u. +37°C at 60% rh
Humidification/dehumidification system:
- active humidification and dehumidification from 10-90% rh with digital display of the relative humidity - display resolution 0.5%, setting accuracy 1%
- Humidity supply with distilled water from external container by means of self-suction pump
- humidification by means of steam generator
- dehumidification by means of cooling trap in Peltier-technology
Multiple overheating protection:
- with acoustic and optical alarm in case of excess or insufficient temperature, warning when water container is empty
- separated electronic temperature selection guard TWW, protection class 3.3, can be digitally set
- additionally integrated protection from excess and insufficient temperature "ASF', which automatically follows the target value in freely selectable distance
Structure stainless-steel housing:
- fully insulated stainless-steel door with double lock and 4-point adjustment
- inner glass door
- fully galvanised back wall made of steel panel
Interior:
- easy-to-handle stainless-steel workspace enforced by deep-drawing ribbing W.-St. 1.4301 (ASTM 304) hermetically welded
- 2 grids made of stainless steel
Heating and cooling system:
- energy-saving Peltier heating/cooling system integrated in back wall
Temperature range:
- from +5°C to +70°C
- temperature deviation at 10°C and 37°C: < ± 0.4°C (HPP 108), ≤ ± 0,5°C (HPP 749)
Power:
10 LED light strips arranged on the side wall of the interior, illumination strength (measured in 10 cm distance) 10,000 LUX (180 µmol/(m²*s), programme-controlled dimming from 0 to 100% in 10 % steps, ramp programming in combination with temperature and humidity
| Article | Order No. | |
| additional stainless steel grid for HPP 108 | 10 26 04038 | |
| additional stainless steel grid for HPP 749 | 10 26 04039 | |
| Stainless steel shelf, perforated, for HPP 108 | 10 26 04020 | |
| Stainless steel shelf, perforated, for HPP 749 | 10 26 04029 | |
| IQ checklist | 41 26 11011 | |
| with works test data as support for validation by customer. | |
| OQ checklist | 10 26 01024 | |
| with cabinet-related factory data, for one freely selectable humidity and temperature value; includes temperature distribution measurement for 27 measuring points pursuant to DIN 12880:2007-05 as support for validation at the customer side | |
| Celsius FDA-Edition | 41 26 11013 | |
| Software conforming to FDA (no domain network support) | |
| Oven-related user authorization card | 41 26 11021 | |
| prevents undesired manipulation by unauthorized third parties | |
| Ethernet interface instead of USB | 10 26 04024 | |
| incl. Software "Celsius Ethernet Edition" | |
| RS232 interface instead of USB | 10 26 04021 | |
| RS485 interface instead of USB | 10 26 04022 | |
| Temperature profile write/read unit | 10 26 04014 | |
| for programming via PC, for writing to and reading from the chip card (up to 40 ramps) | |
| Additional chip card, blank, formatted | 10 26 04015 | |
| 32 kb MEMoryCard XL for a maximum of 40 ramps | |
| USB connection cable for computer interface | 10 26 04023 | |
Special versions like lockable door, duct, additional Pt100s etc. are available on request.