Chemic ALS - Via S.Francesco d'Assisi, 200 - 70018 Rutigliano (BA) - Italy - tel. (+39)080.476.10.66 - fax (+39)080.476.11.26 - info@labcatalogo.it - www.labcatalogo.it
Main -> Heating units, refrigerating units -> Temperature controllers -> Climatic chambers, test chambers
Constant Climate Chamber, Memmert
Models HPP

Standard scope of delivery

Venting and control systems:

  • mandatory interior air circulation by means of circulation ventilators
  • adaptive, fuzzy-controlled, multi-functional and digital PID-micro-processor control, for adjustable control of the Peltier heating/cooling system without cooling agents
  • self-diagnostics system to find errors for temperature and humidity control
  • 2 Pt100-sensors DIN-class A in 4-conductor measurement with changing monitoring and function adaption at exactly the same temperature
  • digital real-time week-programme clock with day-group function, can be adjusted by the minute
  • relative-time countdown clock for tempering profiles with up to 40 ramps, each segment can be set from 1 min to 999 hrs.
  • digital display of all temperature, weekday, time, humidity and set-up default values - language selection via set-up
  • digital target value specifications / current value display: 0.1°C
  • integrated GLP-conform long-term logging (ring buffer) of all relevant data as data logging - 1024 kB
  • saving the programme run in case of power failure
  • parallel printer interface (incl. real-time clock with date function) for all PCL3-compatible ink jet printers for GLP-conform documentation
  • USB interface including Memmert programming and logging software "Celsius" with advanced features
  • chip-card control including 1 MEMoryCard XL with 32 kB memory (max. 40 ramps)
  • incl. factory calibration certificate for +10°C u. +37°C at 60% rh

Humidification/dehumidification system:

  • active humidification and dehumidification from 10-90% rh with digital display of the relative humidity - display resolution 0.5%, setting accuracy 1%
  • Humidity supply with distilled water from external container by means of self-suction pump
  • humidification by means of steam generator
  • dehumidification by means of cooling trap in Peltier-technology

Multiple overheating protection:

  • with acoustic and optical alarm in case of excess or insufficient temperature, warning when water container is empty
  • separated electronic temperature selection guard TWW, protection class 3.3, can be digitally set
  • additionally integrated protection from excess and insufficient temperature "ASF', which automatically follows the target value in freely selectable distance

Structure stainless-steel housing:

  • fully insulated stainless-steel door with double lock and 4-point adjustment
  • inner glass door
  • fully galvanised back wall made of steel panel

Interior:

  • easy-to-handle stainless-steel workspace enforced by deep-drawing ribbing W.-St. 1.4301 (ASTM 304) hermetically welded
  • 2 grids made of stainless steel

Heating and cooling system:

  • energy-saving Peltier heating/cooling system integrated in back wall

Temperature range:

  • from +5°C to +70°C
  • temperature deviation at 10°C and 37°C: < ± 0.4°C (HPP 108), ≤ ± 0,5°C (HPP 749)

Power:

  • 230 V ± 10%), 50/60 Hz
TypeOuter dimensions W × H × D
mm
Inner dimensions W × H × D
mm
Vol. approx.
l
Wght approx.
kg
P rating
W
Order No.
(1)HPP 108710×760×640560×480×400108 66 35010 26 02108
(2)HPP 7491190×1620×8401040×1200×600749218105010 26 02749


Light modules for HPP 108

10 LED light strips arranged on the side wall of the interior, illumination strength (measured in 10 cm distance) 10,000 LUX (180 µmol/(m²*s), programme-controlled dimming from 0 to 100% in 10 % steps, ramp programming in combination with temperature and humidity

ArticleOrder No.
Cold white light module 5,500 Kelvin10 26 01087
Cold white light module 5,500 Kelvin + warm white 2,700 Kelvin10 26 01088


Special equipment
ArticleOrder No.
additional stainless steel grid for HPP 10810 26 04038
additional stainless steel grid for HPP 74910 26 04039
Stainless steel shelf, perforated, for HPP 10810 26 04020
Stainless steel shelf, perforated, for HPP 74910 26 04029
IQ checklist41 26 11011

with works test data as support for validation by customer.

OQ checklist10 26 01024

with cabinet-related factory data, for one freely selectable humidity and temperature value; includes temperature distribution measurement for 27 measuring points pursuant to DIN 12880:2007-05 as support for validation at the customer side

Celsius FDA-Edition41 26 11013

Software conforming to FDA (no domain network support)

Oven-related user authorization card41 26 11021

prevents undesired manipulation by unauthorized third parties

Ethernet interface instead of USB10 26 04024

incl. Software "Celsius Ethernet Edition"

RS232 interface instead of USB10 26 04021
RS485 interface instead of USB10 26 04022
Temperature profile write/read unit10 26 04014

for programming via PC, for writing to and reading from the chip card (up to 40 ramps)

Additional chip card, blank, formatted10 26 04015

32 kb MEMoryCard XL for a maximum of 40 ramps

USB connection cable for computer interface10 26 04023



Special versions like lockable door, duct, additional Pt100s etc. are available on request.