Chemic ALS - Via S.Francesco d'Assisi, 200 - 70018 Rutigliano (BA) - Italy - tel. (+39)080.476.10.66 - fax (+39)080.476.11.26 - info@labcatalogo.it - www.labcatalogo.it
Main -> Heating units, refrigerating units -> Temperature controllers -> Incubators, cooled incubators
Cooled Incubators / Storage Chambers, Memmert

Series ICP compressor cooled
Temperature range: 0°C to +60°C (ICP 400 + 500)
Temperature range: -12°C to +60°C (ICP 600 – 800)
Max. temperature deviations precision ±0.1°C, uniformity (acc. to DIN) at 37°C ±0.3°C.
Turbulence-free internal air-ventilation, separation of the thermal jacket from the interior prevents the occurrence of drying out the loads through the cooling-device. Energy saving heating/cooling system (R134a), highly efficient automatic defrosting system. Segment-dependent control of the internal air-ventilation in 10% steps (from 10 to 100%). 2 stainless steel grids are supplied.

Series IPP Peltier-cooled
Temperature range: +5°C to +70°C
Max. temperature deviations precision better than 0.1°C, uniformity better than ±0.4°C (at +10 and +37°C)
Forced ventilation by Peltier fan. Energy-saving, refrigerant-free Peltier heating/cooling system (heat-pump principle). 1 stainless steel grid each is supplied with the model sizes 200 and 300; on models of sizes 400 and 500 there are 2 stainless steel grids.

Series IPS Peltier-cooled
Temperature range: +14°C to +45°C
Max. temperature deviations precision better than 0.1°C, uniformity better than ±0.4°C (at +18 and +25°C)
Forced ventilation by Peltier fan. Energy-saving, refrigerant-free Peltier heating/cooling system (heat-pump principle). 2 stainless steel grids are comprised in the delivery.

Basic equipment for ICP / IPP / IPS:
Adaptive, fuzzy-supported multifunctional PID process control for - if required set temperature dependent - multiramp (up to max. 40 segments) temperature profiles with integrated 7-day program timer with real-time clock and segment-specific in program operation (40 ramp segments) settable ramp-timer (relative time) from 1 min. up to 999 h. Menu guidance through 8-digit alphanumeric digital display.
Parallel printer interface (incl. a real-time clock with date function) for all PCL3-compatible ink-jet printers for GLP-conforming documentation. Serial interface USB with software "Celsius" with advanced features.
Multiple overtemperature protection, undertemperature protection, protection class 3.3 and gliding over/undertemperature protection linked automatically to the setpoint (ASF = Auto-Safety Function) at a preselected tolerance margin.
Digital display of all temperature and time values as well as visualizing of further important operating functions. Autodiagnostic system for rapid fault identification. Integrated long-term documenting (ring store) of all relevant data. Chip card control; setpoints and actual temperature recording up to 40 ramps. 32 kB MEMory Card XL. Works calibration certificate (chamber center) for 10°C and 37°C (IPP and ICP). Works calibration certificate (chamber center) for +18°C and +25°C (IPS).
All models bear the CE mark.

 

Technical data:

Interior:

made of stainless steel (No. 1.4301/ASTM 304), deep-drawn.

Housing:

made of textured stainless steel, rear zinc-plated steel.

Double doors:

inner glass, outer stainless steel

Electrical supply:

ICP models 230 V (±10%), 50 Hz
IPP/IPS models 230 V (±10 %), 50/60 Hz
Other voltages to special order.



Series ICP
TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
ICP 400558×486×967400×330×400 53 68 50041 26 42400
(1)ICP 500718×556×1047560×400×480108 87 50041 26 42500
ICP 600958×656×1335800×500×640256144 70041 26 42600
(2)ICP 7001198×656×14951040×500×800416178 75041 26 42700
ICP 8001198×756×18951040×600×1200749227120041 26 42800



External dimensions: depth without door handle, depth of door handle 38 mm
Interior: depth less 30 mm for air duct in the middle of back wall

Series IPP
TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
IPP 200550×490×600400×250×320 32 33 12541 26 44200
IPP 300630×490×600480×250×320 39 36 12541 26 44300
IPP 400550×550×680400×330×400 53 43 17541 26 44400
(3)IPP 500710×620×760560×400×480108 66 35041 26 44500
IPP 8001190×825×16201040×600×1200749218105010 26 04800



External dimensions: depth without door handle, depth of door handle 38 mm
Interior: depth less 10 mm for air duct in the middle of back wall

Series IPS
TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
IPS 7491190×825×16201040×600×120074920035010 26 05749



External dimensions: depth without door handle, depth of door handle 38 mm
Interior: depth less 10 mm for air duct in the middle of back wall

Accessories

Stainless steel grids (standard version)

for typeQty max.Order No.for typeQty max.Order No.
IPP 200 310 26 09402ICP 400 410 26 09303
IPP 300 310 26 09403ICP 500 510 26 09304
IPP 400 410 26 09404ICP 600 710 26 09305
IPP 500 510 26 09505ICP 700 910 26 09306
IPP 8001410 26 09406ICP 8001410 26 09308


Stainless steel shelves, perforated
for typeQty max.Order No.for typeQty max.Order No.
IPP 200 341 26 39202ICP 400 441 26 39103
IPP 300 341 26 39203ICP 500 541 26 39104
IPP 400 441 26 39204ICP 600 741 26 39105
IPP 500 541 26 39205ICP 700 941 26 39106
IPP 8001410 26 09206ICP 8001441 26 39108


Special equipment
ArticleOrder No.
Celsius FDA-Edition41 26 11013

Software conforming to FDA (no domain network support)

(4)Oven-related user authorization card41 26 11021

prevents undesired manipulation by unauthorized third parties

Ethernet interface instead of USB10 26 04024

incl. Software "Celsius Ethernet Edition"

RS232 interface instead of USB10 26 04021
RS485 interface instead of USB10 26 04022


Checklists
ArticleOrder No.
IQ checklist41 26 11011

with works test data as support for validation by customer.

OQ checklist (size 200 and 300)10 26 01016

with cabinet-related factory data, for one freely selectable temperature value; includes temperature distribution measurement for 9 measuring points pursuant to DIN 12880:2007-05 as support for validation at the customer side

OQ checklist (size 400 up to 800 and 749)10 26 01017

with cabinet-related factory data, for one freely selectable temperature value; includes temperature distribution measurement for 27 measuring points pursuant to DIN 12880:2007-05 as support for validation at the customer side




We reserve the right to make changes in technical specifications. Dimensions subject to confirmation.
All performance data are typical mean values for serial devices.
Special equipment as entry ports, Pt100 for external temperature recording, a.s.o. upon request.