Chemic ALS - Via S.Francesco d'Assisi, 200 - 70018 Rutigliano (BA) - Italy - tel. (+39)080.476.10.66 - fax (+39)080.476.11.26 - info@labcatalogo.it - www.labcatalogo.it
Main -> Heating units, refrigerating units -> Temperature controllers -> Incubators, cooled incubators
Precision Incubators, Memmert

BASIC:
with microprocessor PID temperature controller with integral autodiagnostic system for rapid fault identification, 1 high-grade Pt100 sensor DIN Class A in 4-wire-circuit, digital timer to switch off heating (1 min. up to 99 h 59 min), incl. digital LED-display of set and actual temperature as well as residual time; double overtemperature protection - in case of a faulty sensor or the total breakdown of the controller the heating is switched off approx. 3°C above the set temperature and additionally at approx. 10°C above the max. oven temperature by the mechanical temperature limiter TB protection class 1.

EXCELLENT:
with fuzzy-supported microprocessor PID controller with integral autodiagnostic system for rapid fault identification, 2 high-grade Pt100 sensors DIN Class A in 4-wire circuit, mutually monitoring and taking over the performance, digital 7-day-program timer with real time clock with group function (e.g. weekdays) for one set value or in combination with the integrated digital timer (with residual time) for ramp operating with max. 4 segments: DELAYED ON, HEATING UP, HOLD or HOLD set-temperature-dependent (guaranteeing the hold period), COOLING DOWN - each segment adjustable from 1 min to 999 hours. Integrated long-term documenting (ring store) of all relevant data, GLP-conforming as data logger - 1024 kB as well as serial computer interface RS232 with software "Celsius" with advanced features. Triple overtemperature protection: in case of overtemperature due to failure, the heating is switched off at approx. 10°C above the set temperature (fixed value), digitally adjustable electronic overtemperature controller, TWW, protection class 3.1, and mechanical temperature limiter, TB, protection class 1, switching the heating off at approx. 10°C above max. oven temperature. Works calibration certificate for one temperature, 37°C (chamber center).

PERFECT:
adaptive, fuzzy-supported multifunctional digital microprocessor PID-controller and integral autodiagnostic system for rapid fault identification, 2 high-grade Pt100 sensor DIN Class A in 4-wire-circuit, mutually monitoring and taking over the performance, digital 7-day-program timer with real time clock with group function (e.g. weekdays), integrated timer for thermostating profiles of up to 40 ramps, each segment adjustable from 1 min. up to 999 hours (residual time); air flap adjustable in 10% steps, on UF models fan speed adjustable in 10% steps, as well as setpoint dependent operation easily programmable, via set-up) visualizing also all set parameters. Integrated long-term documenting (ring store) of all relevant data, GLP-conforming as data logger - 1024 kB (protected against manipulation); parallel printer interface with real-time clock for all Pcl3-compatible ink-jet printers, serial interface RS232 (upon request - without extra charge - RS485) incl. software "Celsius" with advanced features as well as MEMory Card with 32 kB and STERiCard for automatic incubator chamber sterilization cycle (not for sterilizing the load). Multiple overtemperature protection with audible and visual alarm, digitally adjustable electronic protection with audible and visual alarm, digitally adjustable electronic microprocessor overtemperature controller TWW protection class 3.1 - maximum value for overtemperature - minimum value for undertemperature, additionally adjustable auto-safety-function "ASF" for over-and undertemperature automatically following the set value at a preset tolerance range; audible alarm is activated in case of over- or undertemperature, heating is switched off in case of overtemperature, additional mechanical temperature limiter, TB, protection class 1, switching the heating off at approx. 10°C above max. oven temperature. Work calibration certificate for one temperature, 37°C (chamber center).

 

Technical data:

Interior:

stainless steel, material 1.4301 (ASTM 304), reinforced by deep drawn ribbing and easy to clean.

Housing (external):

Textured stainless steel, rear zinc-plated steel, (INE/INP/IFE/IFP 800 mounted on lockable castors.)

Temperature range:

from 30°C (however, at least 5°C above room temperature, 10°C for models with air turbine) up to 70°C.

Temperature deviations:

precision max. ±0.1°C; uniformity (acc. to DIN) at 37°C max. ±0.6°C

Electrical supply:

230 V (±10%), 50/60 Hz. Other voltages to special order.

Available models and sizes with internal glass door and fully insulated stainless steel external door:



Series INB1)

With natural air circulation and electronic PID controller.

TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
INB 200550×400×600400×250×320 322844041 26 70200
INB 300630×400×600480×250×320 393050041 26 70300
(1)INB 400550×480×680400×330×400 533580041 26 70400
INB 500710×550×760560×400×4801085090041 26 70500


Series INE1)

With natural air circulation and electronic PID control with fuzzy support.

TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
INE 200550×400×600400×250×320 32 28 44041 26 80200
INE 300630×400×600480×250×320 39 30 50041 26 80300
INE 400550×480×680400×330×400 53 35 80041 26 80400
(2)INE 500710×550×760560×400×480108 50 90041 26 80500
INE 550630×650×920480×500×640153 66110010 26 08550
INE 600950×650×920800×500×640256 87160041 26 80600
INE 7001190×650×10801040×500×800416121180041 26 80700
INE 8001190×750×16201040×600×1200749170200041 26 80800


Series INP1)

With natural air circulation and multifunctional PID process controller with fuzzy support.

TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
INP 200550×400×600400×250×320 32 28 44041 26 90200
INP 300630×400×600480×250×320 39 30 50041 26 90300
INP 400550×480×680400×330×400 53 35 80041 26 90400
INP 500710×550×760560×400×480108 50 90041 26 90500
INP 550630×650×920480×500×640153 66110010 26 09550
INP 600950×650×920800×500×640256 87160041 26 90600
INP 7001190×650×10801040×500×800416121180041 26 90700
(3)INP 8001190×750×16201040×600×1200749170200041 26 90800


Series IFE1)

With forced air movement and electronic PID control with fuzzy support.

TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
IFE 400550×480×680400×330×400 53 35 80010 26 03401
IFE 500710×550×920560×400×480108 50 90010 26 03501
IFE 550630×650×920480×500×640153 66110010 26 03551
IFE 600950×650×920800×500×640256 87160010 26 03601
IFE 7001190×650×10801040×500×800416121180010 26 03701
IFE 8001190×750×16201040×600×1200749190200010 26 03801


Series IFP1)

With forced air movement and multi-functional PID process control with fuzzy support.

TypeOuter W × D × H
mm
Inner W × D × H
mm
Vol. approx.
l
Wght approx.
kg
P rating approx.
W
Order No.
IFP 400550×480×680400×330×400 53 35 80010 26 02401
IFP 500710×550×760560×400×480108 50 90010 26 02501
IFP 550630×650×920480×500×640153 66110010 26 02551
IFP 600950×650×920800×500×640256 87160010 26 02601
IFP 7001190×650×10801040×500×800416121180010 26 02701
IFP 8001190×750×16201040×600×1200749170200010 26 02801



1) External dimensions: depth without door handle, depth of door handle 38 mm.
All Precision Incubators according to DIN 12880:2007-05. Standard models bear the test marks: EMV/VDE/GS/CE/GOST

Accessories for models BASIC-, EXCELLENT- and PERFECT

Stainless steel grid (standard version)

for typeQty max.Order No.for typeQty max.Order No.
200 310 26 00320550 710 26 00355
300 310 26 00330600 710 26 00360
400 410 26 00340700 910 26 00370
500 510 26 003508001410 26 00380


Stainless steel shelves, perforated

Stainless steel shelves, perforated

for typeQty max.Order No.for typeQty max.Order No.
200 341 26 00020550 710 26 00055
300 341 26 00030600 741 26 00060
400 441 26 00040700 941 26 00070
500 541 26 000508001441 26 00080


Stainless steel shelves

Stainless steel trays, non-perforated, 15 mm rim

for typeQty max.Order No.for typeQty max.Order No.
200 341 26 10020550 710 26 01055
300 341 26 10030600 741 26 10060
400 441 26 10040700 941 26 10070
500 541 26 100508001441 26 10080


Special equipment only for models EXCELLENT- and PERFECT
ArticleOrder No.
IQ checklist41 26 11011

with works test data as support for validation by customer.

OQ checklist (size 200 and 300)10 26 01016

with cabinet-related factory data, for one freely selectable temperature value; includes temperature distribution measurement for 9 measuring points pursuant to DIN 12880:2007-05 as support for validation at the customer side

OQ checklist (size 400 up to 800 and 749)10 26 01017

with cabinet-related factory data, for one freely selectable temperature value; includes temperature distribution measurement for 27 measuring points pursuant to DIN 12880:2007-05 as support for validation at the customer side

USB interface instead of RS23210 26 01014


Special equipment only for model PERFECT
ArticleOrder No.
Celsius FDA-Edition41 26 11013

Software conforming to FDA (no domain network support)

(4)Oven-related user authorization card41 26 11021

prevents undesired manipulation by unauthorized third parties

Ethernet interface instead of RS23210 26 01015

incl. software "Celsius Ethernet Edition"




Further special equipment for EXCELLENT and PERFECT Class, such as entry ports, Pt100 for external temperature recording, a.s.o. upon request.
We reserve the right to make changes in technical specifications. Dimensions subject to confirmation.